Electrical Specifi cations
Specifi cations typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifi cations subject to change without notice.
Key Features: Input
Parameter Conditions Min. Typ. Max. Units
• 2W Output Power
5 VDC Input 4.5 5.0 5.5
• 4 kVAC rms Isolation Input Voltage Range 12 VDC Input 10.8 12.0 13.2 VDC
• Reinforced Insulation 24 VDC Input 21.6 24.0 26.4
Input Filter Internal Capacitor
• Meets 1xMOPP & 2xMOOP
Output
• 2 µA Leakage Current Max Parameter Conditions Min. Typ. Max. Units
• Compact MiniDIP Case Output Voltage Accuracy ±2.0 ±4.0 %
Output Voltage Balance Dual Outputs, Balanced Loads ±0.1 ±1.0 %
• Single & Dual Outputs
Line Regulation VIN = Min to Max ±1.2 ±1.5 %
• 2.0 MH MTBF
Load Regulation, See Note 2 See Model Selection Guide
• Industry Standard Pin-Out Ripple & Noise (20 MHz) See Note 3 100 150 mV P - P
• EN 60601 Approved Temperature Coeffi cient ±0.01 ±0.02 %/°C
Output Short Circuit 0.5 Sec. Max (Autorecovery)
General
Parameter Conditions Min. Typ. Max. Units
Isolation Voltage, Rated 60 Seconds 4,000 VAC rms
Reinforced Insulation Working Voltage 300 Vrms
Leakage Current 2.0 µA
Isolation Resistance 500 VDC 10 G
Isolation Capacitance 100 kHz, 1V 15 20 pF
Switching Frequency 50 80 100 kHz
Conducted EMI Complies With EN 55022, Class A
th
EMC Complies With EN 55011 4 Edition
EMS Complies With EN 60601-1-2
Environmental
Parameter Conditions Min. Typ. Max. Units
Ambient -25 +80 °C
Operating Temperature Range
Case +105 °C
Storage Temperature Range -50 +125 °C
Cooling Free Air Convection
Humidity RH, Non-condensing 95 %
Physical
Case Size See Mechanical Diagram (Page 2)
Case Material Non-Conductive Black Plastic (UL94-V0)
Weight 0.18 Oz (5.1g)
MicroPower Direct Reliability Specifi cations
292 Page Street Parameter Conditions Min. Typ. Max. Units
Suite D MTBF MIL HDBK 217F, 25°C, Gnd Benign 2.0 MHours
rd
Stoughton, MA 02072 UL 60601-1, EN 60601-1 3 Edition
USA Safety Approvals UL 60950-1, EN 60950-1
ANSI/AAMI ES 60601-1 1xMOPP & 2XMOOP Recognition
T: (781) 344-8226 Absolute Maximum Ratings
F: (781) 344-8481 Parameter Conditions Min. Typ. Max. Units
E: sales@micropowerdirect.com
W: www.micropowerdirect.com 5 VDC Input 9.0
Input Voltage Surge (1 Sec) 12 VDC Input 18.0 VDC
24 VDC Input 30.0
Lead Temperature 1.5 mm From Case For 10 Sec 260 °C
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.