Page 1 - MPD - MD100MHI Datasheet
P. 1





MD100MHI Series



Miniature, 1W SIP

Medical Approved

DC/DC Con vert ers



Electrical Specifi cations
Specifi cations typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifi cations subject to change without notice.
Key Features: Input
Parameter Conditions Min. Typ. Max. Units
• 1W Output Power
5 VDC Input 4.5 5.0 5.5
• 3 kVAC rms Isolation Input Voltage Range VDC
12 VDC Input 10.8 12.0 13.2
• Reinforced Insulation Input Filter LC Filter
• Meets 1xMOPP & 2xMOOP Output
• Miniature SIP Case Parameter Conditions Min. Typ. Max. Units
• Single & Dual Outputs Output Voltage Accuracy ±1.0 ±3.0 %
• 12 Standard Models Output Voltage Balance Dual Output , Balanced Loads ±0.1 ±1.0 %
• 2.0 MH MTBF Line Regulation For VIN Change of 1% ±1.2 ±1.5 %
• Industry Standard Pin-Out Load Regulation, See Note 2 See Model Selection Guide
• EN 60601 Approved Ripple & Noise (20 MHz) See Note 3 150 mV P - P
Temperature Coeffi cient ±0.01 ±0.02 %/°C
Output Short Circuit Momentary (0.5 Sec.)
General
Parameter Conditions Min. Typ. Max. Units
Isolation Voltage 60 Seconds 3,000 VAC rms
Reinforced Insulation Working Voltage 300 Vrms
Isolation Resistance 500 VDC 10 G
Isolation Capacitance 100 kHz, 1V 15 20 pF
Switching Frequency 50 80 100 kHz
th
EMC Complies With EN 55011 4 Edition
EMS Complies With EN 60601-1-2
Environmental
Parameter Conditions Min. Typ. Max. Units
Ambient -25 +85 °C
Operating Temperature Range
Case +90 °C
Storage Temperature Range -50 +125 °C
Cooling Free Air Convection
Humidity RH, Non-condensing 95 %
Physical
Case Size See Mechanical Diagram (Page 2)
Case Material Non-Conductive Black Plastic (UL94-V0)
MicroPower Direct Weight 0.13 Oz (3.9g)
292 Page Street Reliability Specifi cations
Suite D Parameter Conditions Min. Typ. Max. Units
Stoughton, MA 02072
USA MTBF MIL HDBK 217F, 25°C, Gnd Benign 2.0 MHours
UL 60601-1, EN 60601-1 3rd Edition, UL 60950, EN 60950
Safety Approvals
T: (781) 344-8226 ANSI/AAMI ES 60601-1 1xMOPP & 2XMOOP Recognition
F: (781) 344-8481 Absolute Maximum Ratings
E: sales@micropowerdirect.com
Parameter Conditions Min. Typ. Max. Units
W: www.micropowerdirect.com
5 VDC Input 9.0
Input Voltage Surge (1 Sec) VDC
12 VDC Input 29.0
Lead Temperature 1.5 mm From Case For 10 Sec 260 °C
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com