Page 1 - MPD MPM-40M Datasheet
P. 1





MPM-40M Series



Medical Approved, 40W


Single & Dual Output

AC/DC Power Supplies



Key Features: Electrical Specifi cations
Specifi cations typical @ +25°C, 230 VAC input voltage & rated output current, unless otherwise noted. Specifi cations subject to change without notice.
• 40W Output Power Input
RD
• EN 60601-1 3 Edition Parameter Conditions Min. Typ. Max. Units
85 264 VAC
• Universal 85-264 VAC Input Input Voltage Range
120 370 VDC
• EN 60950 Approved Input Frequency 47 440 Hz
• Reinforced Insulation Inrush Current, See Note 2 115 VAC 30 A Pk
230 VAC 60
• Meets IEC Safety Class II No Load Power Consumption 0.3 W
• Low Leakage Current Output
Parameter Conditions Min. Typ. Max. Units
• Single and Dual Outputs
Output Voltage Accuracy ±2.0 %
• Meets UL 508 Line Regulation VIN = Min to Max ±0.5 %
Single Output ±1.0
Load Regulation, IO = Min to Max %
Dual Output ±2.0
Chassis Mount Models Ripple & Noise (20 MHz), See Note 3 5.0 VDC Output 1.5 1.8 %VP-P of VO
& DIN Rail Mount All Other Outputs 1.0 1.3
Option Available! Hold-Up Time 115 VAC, 60 Hz 25 mS
230 VAC, 50 Hz 80
Temperature Coeffi cient ±0.02 %/°C
Over Voltage Protection See Note 4 120 % of VO
Overshoot 5%
Overload Protection See Note 5 105 % INOM
Over Temperature Protection See Note 6 90 °C
Short Circuit Protection, See Note 7 Continuous (Autorecovery)
General
Parameter Conditions Min. Typ. Max. Units
Isolation Voltage (Reinforced) Input to Output 4,000 VAC rms
Leakage Current 80 µA
Isolation Resistance 500 VDC 1,000 M
Switching Frequency 130 kHz
Environmental
Parameter Standard Criterea Level
Radiated Emissions EN 55014, EN 55024 Class B
Conducted Emissions EN 55022 Class B
A ±8 kV Air
ESD EN 61000-4-2
A ±4 kV Contact
RS EN 61000-4-3 A 10 V/m
EFT EN 61000-4-4 A ±2 kV
Surge EN 61000-4-5 A ±1 kV
CS EN 61000-4-6 A 10 Vrms
PFM EN 61000-4-8 A 30 A/M
Voltage Dips EN 61000-4-11 A 30% 10 mS
Interruptions EN 61000-4-11 B 95% 5,000 mS
MicroPower Direct
Environmental
292 Page Street Parameter Conditions Min. Typ. Max. Units
Suite D Operating Temperature Range Ambient -40 +25 +80 °C
Stoughton, MA 02072 Storage Temperature Range -40 +95 °C
USA Cooling, See Note 8 Free Air Convection (See Derating Curve)
Humidity RH, Non-condensing 95 %
T: (781) 344-8226 Physical
F: (781) 344-8481 Case Size See Mechanical Diagrams (Pages 3 & 4)
E: sales@micropowerdirect.com Case Material Non-Conductive Plastic Resin (UL94-V0)
W: www.micropowerdirect.com Reliability Specifi cations
Parameter Conditions Min. Typ. Max. Units
MTBF MIL HDBK 217F, 25°C, Gnd Benign 200 kHours
Safety Standards UL 60950, EN 60950
Safety Class IEC 60536 Class II

www.micropowerdirect.com