Page 1 - MPD MPM-05MS Datasheet
P. 1





MPM-05M Series



Compact Encapsulated,

Medical Approved, 5W


AC/DC Power Supplies



Key Features: Electrical Specifi cations
Specifi cations typical @ +25°C, 230 VAC input voltage & rated output current, unless otherwise noted. Specifi cations subject to change without notice.
• 5W Output Power
Input
• EN 60601-1 3 Edition
RD
Parameter Conditions Min. Typ. Max. Units
• Universal 85-264 VAC Input 85 264 VAC
Input Voltage Range
• Reinforced Insulation (2xMOPP) 100 370 VDC
• Meets IEC Safety Class II Input Frequency 47 63 Hz
• Short Circuit Protection 115 VAC 10
Inrush Current, See Note 2 A Pk
• Over Current Protection 230 VAC 20
• Over Voltage Protection No Load Power Consumption 0.3 W
• Low Leakage Current Output
Parameter Conditions Min. Typ. Max. Units
• -25°C to +70°C Operation
Output Voltage Accuracy ±2.0 %
Line Regulation VIN = Min to Max ±0.5 %
Load Regulation IOUT = 10% to 100% ±1.0 %
Ripple & Noise (20 MHz) See Note 3 50 100 mV Pk-Pk
115 VAC, 60 Hz 10
Hold-Up Time mS
230 VAC, 50 Hz 80
Temperature Coeffi cient ±0.02 %/°C
Overload Protection Autorecovery 110 280 % INOM
Short Circuit Protection Continuous (Autorecovery)
General
Parameter Conditions Min. Typ. Max. Units
Isolation Voltage Input to Output, 60S 4,000 VAC rms
Leakage Current 80 µA
Isolation Resistance 500 VDC 1,000 M
Switching Frequency 140 kHz
Environmental
Parameter Conditions Min. Typ. Max. Units
Operating Temperature Range Ambient -25 +25 +70 °C
Maximum Case Temperature Case +95 °C
Storage Temperature Range -40 +95 °C
Cooling Free Air Convection (See Derating Curve)
Humidity RH, Non-condensing 95 %RH
MicroPower Direct Lead Temperature See Note 4 260 °C
292 Page Street 360
Suite D Physical
Stoughton, MA 02072 Case Size See Mechanical Diagram (Page 4)
USA
Case Material Non-Conductive Plastic Resin (UL94-V0)
T: (781) 344-8226 Weight 1.52 Oz (43g)
F: (781) 344-8481 Reliability Specifi cations
E: sales@micropowerdirect.com
W: www.micropowerdirect.com Parameter Conditions Min. Typ. Max. Units
MTBF MIL HDBK 217F, 25°C, Gnd Benign 300 kHours
Safety Standards UL 60601, EN 60601
Safety Class Class II
Insulation Level 2XMOPP
www.micropowerdirect.com
   1   2   3   4